Spectrum Designs Inc Spectrum Designs Medical Anatomical I (Silhouette) Malar Facial Implant, Catalog Number 120-224S. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Spectrum Designs Medical Anatomical I (Silhouette) Malar Facial Implant, Catalog Number 120-224S.
Brand
Spectrum Designs Inc
Lot Codes / Batch Numbers
Lot number 30838, serial numbers BAE256 through BAE275, manufactured June, 2006. Expiration date is five years from date of manufacture.
Products Sold
Lot number 30838, serial numbers BAE256 through BAE275, manufactured June, 2006. Expiration date is five years from date of manufacture.
Spectrum Designs Inc is recalling Spectrum Designs Medical Anatomical I (Silhouette) Malar Facial Implant, Catalog Number 120-224S. due to The product is being recalled because it was discovered the at least one unit in the lot contained a mismatched pair (two rights and no left). Since t. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The product is being recalled because it was discovered the at least one unit in the lot contained a mismatched pair (two rights and no left). Since the lot was produced with equal numbers of pairs, it is assumed that one other unit was packaged with two lefts and no right. This was discovered through a physician complaint.
Recommended Action
Per FDA guidance
The firm initiated the recall on 06/15/2007 by contacting their customers by telephone. The firm followed-up by sending recall letters via FedEx or certified mail with signature confirmation to each direct account who received the recalled product on 06/18/2007.