Steris Corp ASC2000 Ambulatory Surgical Table. Model #MTS110003. The device is used to provide complete and flexible positioning for surgical procedures. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ASC2000 Ambulatory Surgical Table. Model #MTS110003. The device is used to provide complete and flexible positioning for surgical procedures.
Brand
Steris Corp
Lot Codes / Batch Numbers
All serial numbers.
Products Sold
All serial numbers.
Steris Corp is recalling ASC2000 Ambulatory Surgical Table. Model #MTS110003. The device is used to provide complete and f due to Upon investigation, the firm determined that repetitive rubbing of the hydraulic hoses against the interior of the stainless steel shroud may result i. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Upon investigation, the firm determined that repetitive rubbing of the hydraulic hoses against the interior of the stainless steel shroud may result in the rupture of the hydraulic hoses causing loss of control of the table surface.
Recommended Action
Per FDA guidance
A Field Correction Notice dated April 17, 2009 was issued to customers. The letter described the potential problem of the unit. The letter also notified customers of the corrective actions to be taken by the firm. A Steris technician will visit affected accounts to inspect the hydraulic hoses and replace any that show abrasion and will make other preventive corrections. For further information or if you have questions regarding the servicing of your unit, please contact the STERIS Field Service Dispatch at 1-800-288-2805.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MD, MI, MN, MO, NH, NJ, NY, NC, OH, OK, OR, PA, SC, TN, TX, VA, WV, PR