Steris Corp Steris QFC1728 and QFC1729 Quick Connects used in connections and processing endoscopes. The QFC1728 is used to process the Fujinon 250/270/450/470/490 Group 5 and 530/590 series GI Endoscopes w/o Water Jet. The QFC1729 is used to process 250/450 Group 5 and 530/590 series Endoscopes w/ Water Jet. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Steris QFC1728 and QFC1729 Quick Connects used in connections and processing endoscopes. The QFC1728 is used to process the Fujinon 250/270/450/470/490 Group 5 and 530/590 series GI Endoscopes w/o Water Jet. The QFC1729 is used to process 250/450 Group 5 and 530/590 series Endoscopes w/ Water Jet.
Brand
Steris Corp
Lot Codes / Batch Numbers
Lot numbers: 2914737, 2876316, 2871200, 2876274, 2914752, 2876704, 2876720, 2876746, 2876761, 3072030, and 3072014.
Products Sold
Lot numbers: 2914737, 2876316, 2871200, 2876274, 2914752, 2876704, 2876720, 2876746, 2876761, 3072030, and 3072014.
Steris Corp is recalling Steris QFC1728 and QFC1729 Quick Connects used in connections and processing endoscopes. The QFC1728 due to The center insert of the adaptor #6 restrictor could become loose and/or come out of the housing and may affect the sterilization process of the attac. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The center insert of the adaptor #6 restrictor could become loose and/or come out of the housing and may affect the sterilization process of the attached endoscope
Recommended Action
Per FDA guidance
On February 22, 2007, customers were contacted by phone and were notified by recall letter dated February 23, 2007. The recall notifications indicate that Steris Account managers will visit each consignee and replace affect product.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, CO, ID, IN, IA, MA, MO, MT, NV, NJ, PA, SC, TN, TX, WY