Olympus Leak Test Adapter (Steris) – Cleaning Solution Risk (2020)
Potential cleaning solution intrusion can compromise endoscope integrity.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Olympus Leak Test Adapter used with Acu-sInQ Complete Endoscope Cleaning Aid Systems-intended to assist manual flexible endoscope cleaning Part number: 0600818OKN
Brand
Steris Corporation
Lot Codes / Batch Numbers
Distributed from May 23, 2017 to June 30, 2020
Products Sold
Distributed from May 23, 2017 to June 30, 2020
Steris Corporation is recalling Olympus Leak Test Adapter used with Acu-sInQ Complete Endoscope Cleaning Aid Systems-intended to ass due to Potential for cleaning solution intrusion into endoscope. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential for cleaning solution intrusion into endoscope
Recommended Action
Per FDA guidance
Steris issued Urgent Medical Device Recall issued letter on 10/7/20 stating reason for recall, health risk and action to take: Please immediately inspect your facility for affected Leak Test Adapters that connect to Olympus endoscopes. The affected Leak Test Adapters have a white plastic valve connector. If you have this Leak Test Adapter, please disconnect from your Acu-sInQ Complete, destroy, and replace with the new Leak Test Adapter included with this Customer Notification Letter. 2. Please complete the Medical Device Recall Response Form included with this Customer Notification Letter. Your STERIS Sales Representative can assist you should you have any questions while completing the form, or when replacing the Leak Test Adapter. 3. Return the completed Medical Device Recall Response Form to STERIS via one of the following three methods: a. Visit https://www.steris.com/medical-device-recall-response-form and complete the form electronically; b. Scan and email the completed form to Regulatory_Compliance@STERIS.com; or c. Fax the completed form to (440) 392-8963. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact Katilin Hawranko, Product Manager at (440) 392-7431, STERIS Customer Service at 1-800-548-4873, or your local STERIS Representative.