Steris Corporation Steris Amsco Sonic Energy Cleaner, Part Nos.: P080000005, B018905020, B018905JBJ, and B018905JBJSS Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Steris Amsco Sonic Energy Cleaner, Part Nos.: P080000005, B018905020, B018905JBJ, and B018905JBJSS
Brand
Steris Corporation
Lot Codes / Batch Numbers
All units bearing serial numbers 0432000018 to 0403105051.
Products Sold
All units bearing serial numbers 0432000018 to 0403105051.
Steris Corporation is recalling Steris Amsco Sonic Energy Cleaner, Part Nos.: P080000005, B018905020, B018905JBJ, and B018905JBJSS due to Smoke, sparking and fire hazard-Units manufactured between November 2000 and January 2005 are susceptible to water damage at the lid switch cable wiri. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Smoke, sparking and fire hazard-Units manufactured between November 2000 and January 2005 are susceptible to water damage at the lid switch cable wiring harness, which may result in smoking, sparking, and fire.
Recommended Action
Per FDA guidance
The firm sent out a certified letter to all affected consignees alerting them to problems with the device on 6/22/06.