Stille AB Gardsvagen Stille Sonesta Gynecology and Urology Table 6302 and 6302U; an electric cystometric table; Stille AB, Solna, Sweden. Used to position the subject/patient in a desired position during urodynamic, general examinations, cystoscopy, ultrasound, biofeedback, vasectomy and OBGYN procedures. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Stille Sonesta Gynecology and Urology Table 6302 and 6302U; an electric cystometric table; Stille AB, Solna, Sweden. Used to position the subject/patient in a desired position during urodynamic, general examinations, cystoscopy, ultrasound, biofeedback, vasectomy and OBGYN procedures.
Brand
Stille AB Gardsvagen
Lot Codes / Batch Numbers
Model 6302, serial numbers 12-11075 and Model 6302U, serial numbers 12-11068, 12-11069, 12-11070, 12-11071, 12-11072, 12-11073, 12-11074, 12-11128, 12-11130, 12-11131, 12-11135, 12-11136, 12-11138, 12-11139, 12-11143, 12-11149, tables sold between 3/24/09 and 9/30/09
Products Sold
Model 6302, serial numbers 12-11075 and Model 6302U, serial numbers 12-11068, 12-11069, 12-11070, 12-11071, 12-11072, 12-11073, 12-11074, 12-11128, 12-11130, 12-11131, 12-11135, 12-11136, 12-11138, 12-11139, 12-11143, 12-11149; tables sold between 3/24/09 and 9/30/09
Stille AB Gardsvagen is recalling Stille Sonesta Gynecology and Urology Table 6302 and 6302U; an electric cystometric table; Stille AB due to There is a potential hazard to the patient if the bolts holding the side rail break, causing the accessories attached at the side rails of the tables . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a potential hazard to the patient if the bolts holding the side rail break, causing the accessories attached at the side rails of the tables and chairs to fall off during a medical procedure.
Recommended Action
Per FDA guidance
Stille sent Field Safety Notices dated 10/30/09 to the direct accounts, informing them that there is a potential hazard with the Sonesta tables if the bolts holding the side rail break, causing the accessories attached at the side rails of the tables and chairs to fall off during a medical procedure. Stille developed a retrofit kit consisting of six side rail bolts and instructions for installing the bolts, which were sent to the end users with the safety notice. The accounts were requested to replace the side rail bolts with the new bolts provided, and to complete and return the enclosed acknowledgement card to Stille by fax or e-mail as soon as possible and no later than 12/11/09. Contact Stille Surgical Inc. at 224-612-5404 for assistance.