Stockinette, Single Ply Econo Weave, 8 X 72'', Ref #9999-72. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Stockinette, Single Ply Econo Weave, 8 X 72'', Ref #9999-72.
Lot Codes / Batch Numbers
Lot: 1503929
Products Sold
Lot: 1503929
A medical device manufacturer is recalling Stockinette, Single Ply Econo Weave, 8 X 72'', Ref #9999-72. due to Surgical drapes and covers were packaged in containers with incomplete seals which compromise product sterility.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Surgical drapes and covers were packaged in containers with incomplete seals which compromise product sterility.
Recommended Action
Per FDA guidance
The recalling firm ceased distribution of Stockinette products on 8/26/04 and subsequently notified distributors and direct accounts by letter 9/9/04. The notification was product and lot specific and advised consignees of compromised sterility. Consignees were advised to return response form and product for credit. Distributors were asked to forward the recall notice and respond with a consignee list if they had not notified their accounts. Sub accounts and end users were notified by letter 9/20/04.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026