Straumann Usa, Llc Straumann NC Closure Screw, Titanium, Precision attachment, Article Number: 024.2105.04, Straumann, Andover, MA 01810 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Straumann NC Closure Screw, Titanium, Precision attachment, Article Number: 024.2105.04, Straumann, Andover, MA 01810
Brand
Straumann Usa, Llc
Lot Codes / Batch Numbers
Lot Number: G8677
Products Sold
Lot Number: G8677
Straumann Usa, Llc is recalling Straumann NC Closure Screw, Titanium, Precision attachment, Article Number: 024.2105.04, Straumann, due to Mislabeled package: Narrow Connection (NC) Closure Screws contain closure screws for the regular connection (RC) bone level implant. The RC closures . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Mislabeled package: Narrow Connection (NC) Closure Screws contain closure screws for the regular connection (RC) bone level implant. The RC closures screw is larger (3.5mm diameter) and therefore, will not fit into the NC implant (3.0mm diameter) narrow connection.
Recommended Action
Per FDA guidance
Consignees were notified by telephone on 1/10/08 followed by an Urgent Device Recall letter on January 10, 2008. The letter instructs users to return any affected product to the firm along with a recall acknowledgement form. For additional information, contact 978-747-2514.