Stryker Communications Corp VERTIER Surgical Table, Catalog #0788-100-000 (Standard table); distributed by Stryker Communications, Flower Mound, TX. Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
VERTIER Surgical Table, Catalog #0788-100-000 (Standard table); distributed by Stryker Communications, Flower Mound, TX. Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system.
Brand
Stryker Communications Corp
Lot Codes / Batch Numbers
All lot codes/serial numbers.
Products Sold
All lot codes/serial numbers.
Stryker Communications Corp is recalling VERTIER Surgical Table, Catalog #0788-100-000 (Standard table); distributed by Stryker Communication due to Hydraulic lines responsible for tilting the surgical table have the potential to be severed when articulating table to its lowest position, possibly r. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Hydraulic lines responsible for tilting the surgical table have the potential to be severed when articulating table to its lowest position, possibly resulting in the unexpected and rapid movement of the table.
Recommended Action
Per FDA guidance
Urgent: Device Recall letters were sent to consignees on 07/02/08 via FedEx with return receipt. Consignees were asked to discontinue use of the tables until a Stryker representative has come to the facility and updated the table. The firm has identified and validated replacement parts that prevent this occurrence.