Stryker Endoscopy PKG, 3MM INSERT, METZENBAUM, CURVED, LONG SCISSORS, 29CM, P/N 0250282047 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, an Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
PKG, 3MM INSERT, METZENBAUM, CURVED, LONG SCISSORS, 29CM, P/N 0250282047 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, an
Brand
Stryker Endoscopy
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
641233 643821 646578 745840 747065 747483 841599 845879 846472 847843 944821 945742 946006 1040532 1040542 1042101 1042528 1042785 1042786 1042793 1042794 1046582 1141967 1144235 1145631 1240544 1242137 1245020 1344991 1344992 1345226 1346216 1441375 1441649 1442172
Stryker Endoscopy is recalling PKG, 3MM INSERT, METZENBAUM, CURVED, LONG SCISSORS, 29CM, P/N 0250282047 Laparoscopic Manual Instru due to The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrappe. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026