Stryker Endoscopy PKG, ATRAUMATIC FORCEPS, SINGLE ACTION, P/N 0250080310. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating fu Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
PKG, ATRAUMATIC FORCEPS, SINGLE ACTION, P/N 0250080310. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating fu
Brand
Stryker Endoscopy
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
42714 43489 43490 143175 143176 143452 143453 340645 444574 641197 643520 647411 7108579 740242 840911 844136 848107 941058 942653 943254 943440 945124 1040231 1040393 1042255 1042649 1042650 1043453 1046589 1046699 1046906 1141173 1141190 1141384 1143126 1145283 1145682 1145977 1146679 1243100 1243539 1245517 1246171 1340193 1340476 1340586 1341582 1342389 1343704 1344285 1345531 1440912 1441963 1442257 1442810 1443396
Stryker Endoscopy is recalling PKG, ATRAUMATIC FORCEPS, SINGLE ACTION, P/N 0250080310. Laparoscopic Manual Instruments are intende due to The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrappe. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 12, 2026