Stryker Endoscopy PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080361. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functi Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080361. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functi
Brand
Stryker Endoscopy
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
513592 817411 817416 817475 817517 913167 913168 913323 913324 913522 913555 913556 913559 913560 913623 913624 913986 913987 913989 914273 914275 914276 914277 914474 914753 914754 1014636 1014637 1016127 1016731 1016732 1016733 1113017 1113021 1113023 1113183 1114206 1114207 1114208 1114629 1114630 1114631 1114636 1115403 1115404 1115405 1115406 1115928 1115967 1115968 1115972 1116257 1116654 1116655 1116656 1116720 1213970 1213971 1213972 1214400 1214969 1214970 1215122 1216019 1216020 1216021 1216022 1216819 1216820 1216821 1217044 1313484 1313487 1313488 1313935 1313936 1314206 1314207 1314208 1314691 1314692 1314693 1314694 1314927 1314975 1410317
Stryker Endoscopy is recalling PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080361. Laparoscopic Manual Instruments are intended fo due to The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrappe. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 12, 2026