Stryker Endoscopy Stryker Digital Capture System Ultra (SDC Ultra), Model Number 240-050-988, Stryker Endoscopy, image and video archiving system. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Stryker Digital Capture System Ultra (SDC Ultra), Model Number 240-050-988, Stryker Endoscopy, image and video archiving system.
Brand
Stryker Endoscopy
Lot Codes / Batch Numbers
Serial numbers 07H041654 - 08A069464, PP1.04
Products Sold
Serial numbers 07H041654 - 08A069464, PP1.04
Stryker Endoscopy is recalling Stryker Digital Capture System Ultra (SDC Ultra), Model Number 240-050-988, Stryker Endoscopy, image due to Patient Identification numbers and names may be applied to images from another patient due to a software issue. This failure mode can only occur when. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Patient Identification numbers and names may be applied to images from another patient due to a software issue. This failure mode can only occur when the pre-load function of the SDC ultra is used.
Recommended Action
Per FDA guidance
Stryker Endoscopy mailed Urgent Device Correction notices, dated February 14, 2008, to all consignees, and sales representatives will be supplied with a fix to distribute to the user base.