Stryker Howmedica Osteonics Corp. Holding Forceps for Trident Acetabular System; non Sterile; Catalog number 2105-0000; Howmedica Osteonics Corp, 325 Corporate Drive, Mahwah, NJ 07430; Authorized representative in Europe: Stryker France, Cedex. France The Holding Forceps for Acetabular trial cups help facilitate the insertion / correction of the Acetabular trial inserts. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Holding Forceps for Trident Acetabular System; non Sterile; Catalog number 2105-0000; Howmedica Osteonics Corp, 325 Corporate Drive, Mahwah, NJ 07430; Authorized representative in Europe: Stryker France, Cedex. France The Holding Forceps for Acetabular trial cups help facilitate the insertion / correction of the Acetabular trial inserts.
Brand
Stryker Howmedica Osteonics Corp.
Lot Codes / Batch Numbers
All lots beginning with V1 (XXX).
Products Sold
All lots beginning with V1 (XXX).
Stryker Howmedica Osteonics Corp. is recalling Holding Forceps for Trident Acetabular System; non Sterile; Catalog number 2105-0000; Howmedica Oste due to The tips of the forceps were larger than the positioning holes located on the trials and the implant liners.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The tips of the forceps were larger than the positioning holes located on the trials and the implant liners.
Recommended Action
Per FDA guidance
Important Product Correction letters, dated January 24, 2007, were sent to all consignees on via Federal Express. The letter stated the problem and asked customers to examine their inventory and hospital locations to identify the affected product. All affected forceps product should be retrieved from hospitals and returned to their branch or agency warehouse for reconciliation. Customers are also to reconcile all affected product on the attached Product Accountability Form and fax a copy of that form to 201-831-6069 within 5 days of receipt of the letter. All affected forceps product is to be destroyed and discarded at their branch location. Questions should be directed to their Stryker Orthopaedics Customer Service rep at 1-866-OR-ASSIST.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.