Stryker Instruments Div. of Stryker Corporation Stryker 100 mL AutoFuser PainPump, single site continuous, 1.0 mL/hr, 1 x 2.5" ExFen catheter, sterile, REF 0531-100-025. Infusion pump for pain medication. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Stryker 100 mL AutoFuser PainPump, single site continuous, 1.0 mL/hr, 1 x 2.5" ExFen catheter, sterile, REF 0531-100-025. Infusion pump for pain medication.
Brand
Stryker Instruments Div. of Stryker Corporation
Lot Codes / Batch Numbers
Lot 09112926839.
Products Sold
Lot 09112926839.
Stryker Instruments Div. of Stryker Corporation is recalling Stryker 100 mL AutoFuser PainPump, single site continuous, 1.0 mL/hr, 1 x 2.5" ExFen catheter, steri due to Failure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Failure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
Recommended Action
Per FDA guidance
Stryker sent all consignees a Medical Device Recall Notification letter dated 5/14/09 and instructed to locate and destroy the recalled products and to complete and return the business reply card.