Stryker Medical Div. of Stryker Corporation Stryker Medical Stair-PRO Stair-Chair with wheels, patient transport device, Model 6251. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Stryker Medical Stair-PRO Stair-Chair with wheels, patient transport device, Model 6251.
Brand
Stryker Medical Div. of Stryker Corporation
Lot Codes / Batch Numbers
Serial numbers 080940844, 080940845, 080940858, 080940819, 080940820 and 080940640.
Products Sold
Serial numbers 080940844, 080940845, 080940858, 080940819, 080940820 and 080940640.
Stryker Medical Div. of Stryker Corporation is recalling Stryker Medical Stair-PRO Stair-Chair with wheels, patient transport device, Model 6251. due to Handle could come off. The bolts may back out over time, resulting in a risk of dropping the patient and/or injury to transport personnel, a delay i. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Handle could come off. The bolts may back out over time, resulting in a risk of dropping the patient and/or injury to transport personnel, a delay in medical treatment, or a delay in evacuation.
Recommended Action
Per FDA guidance
Consignees were notified by Urgent Medical Device Correction letter dated 11/18/08 and given the serial numbers of their units, an upgrade kit and instructions for performing replacement of the bolts. Consignees were informed Stryker will perform the upgrade if requested.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026