Trevo NXT ProVue (Stryker) – Endotoxin Testing (2024)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Trevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE, Catalog: 91412 TREVO NXT 4X28+TRAK21 2-PACK - CE, Catalog: 91413 TREVO NXT 4X41+TRAK21 2-PACK - CE, Catalog: 91414 TREVO NXT 6X37+TRAK21 2-PACK - CE, Catalog: 91415 TREVO NXT 3X32+PRO14 2-PACK - CE, Catalog: 91416 TREVO NXT 4X28+TRAK21+CAT6 3-PACK - CE, Catalog: SFP914170 TREVO NXT 6X37+TRAK21+CAT6 3-PACK - CE, Catalog: SFP914190
Brand
Stryker Neurovascular
Lot Codes / Batch Numbers
Catalog/UDI-DI/Lot: 90312/07613327299922/118881, 91412/07613327500684/48483279, 48483281, 48483294, 48696618, 48696619, 48696620, 48715713, 48834692, 48834695, 48834696, 91413/07613327500691/48483280, 48483284, 48483289, 48483291, 48483316, 48483317, 48696613, 48696622, 48696623, 48715711, 48715715, 91414/07613327500707/48447776, 48447778, 48447779, 48483290, 48483292, 48483293, 48483315, 48483318, 48483320, 48696624, 48715714, 48834690, 48834691, 48834694, 48834697, 48852726, 48852730, 48852731, 48852732, 48852734, 48852735, 48852737, 48852738, 91415/07613327500714/48483277, 48483283, 48483295, 48483296, 48483311, 48483312, 48696621, 48715712, 48715716, 48834688, 48834693, 48834698, 91416/07613327507904/47328863, 47328865, 47351815, 47351816, 47374814, 48500988, 48501006, 48501007, 48501008, 48501009, 48501010, 48501011, 48501012, 48501013, 48501014, 48540277, 48556780, 48556783, 48558375, 48588482, 48589876, 48641524, 50065509, 50065520, 50193572, 50260581, 50327385, 50327403, 50375186, 50375187, 50422577, 50422588, 50448274, 50738892, 50762691, 50762721, 50762722, 50786687, 50797491, 50797494, 50797495, 50886508, 51320490, 51320491, 51379394, 51379399, 51379400, 51379401, 51379402, 51379403, 51379404, 51379405, 51379406, 51379408, 51467927, 51489221, 51489230, 51489233, 51489234, SFP914170/07613327508055/48834689, SFP914190/07613327508079/48945914
Products Sold
Catalog/UDI-DI/Lot: 90312/07613327299922/118881; 91412/07613327500684/48483279, 48483281, 48483294, 48696618, 48696619, 48696620, 48715713, 48834692, 48834695, 48834696; 91413/07613327500691/48483280, 48483284, 48483289, 48483291, 48483316, 48483317, 48696613, 48696622, 48696623, 48715711, 48715715; 91414/07613327500707/48447776, 48447778, 48447779, 48483290, 48483292, 48483293, 48483315, 48483318, 48483320, 48696624, 48715714, 48834690, 48834691, 48834694, 48834697, 48852726, 48852730, 48852731, 48852732, 48852734, 48852735, 48852737, 48852738; 91415/07613327500714/48483277, 48483283, 48483295, 48483296, 48483311, 48483312, 48696621, 48715712, 48715716, 48834688, 48834693, 48834698; 91416/07613327507904/47328863, 47328865, 47351815, 47351816, 47374814, 48500988, 48501006, 48501007, 48501008, 48501009, 48501010, 48501011, 48501012, 48501013, 48501014, 48540277, 48556780, 48556783, 48558375, 48588482, 48589876, 48641524, 50065509, 50065520, 50193572, 50260581, 50327385, 50327403, 50375186, 50375187, 50422577, 50422588, 50448274, 50738892, 50762691, 50762721, 50762722, 50786687, 50797491, 50797494, 50797495, 50886508, 51320490, 51320491, 51379394, 51379399, 51379400, 51379401, 51379402, 51379403, 51379404, 51379405, 51379406, 51379408, 51467927, 51489221, 51489230, 51489233, 51489234; SFP914170/07613327508055/48834689; SFP914190/07613327508079/48945914
Stryker Neurovascular is recalling Trevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE, Catalog: 91412 due to Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance wit. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
Recommended Action
Per FDA guidance
On 4/23/24, recall notices were mailed to Supply Chain Management, Recall Coordinators, and Inventory Managers who were asked to do the following: 1) Cease further distribution or use of all affected devices. 2) Segregate the affected units in a secure location for return to the firm. 3) Circulate this Recall-Removal notice internally to all interested/affected parties. 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. 5) Inform the firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so the firm can inform the recipients appropriately. 6) Inform the firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: NVCustomerCare@stryker.com) 7) Complete and return the response form via email to nvfieldactions@stryker.com