Sunrise Medical Inc Manual Wheelchairs with Transit System Option, Model: Quickie LXI (EILXI), Product is manufactured and distributed by Sunrise Medical, Inc. Fresno, CA Intended to provide mobility to persons restricted to a sitting position and are unable to transfer from their wheelchairs when traveling in motor vehicles, the wheelchair must serve as a vehicle seat. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Manual Wheelchairs with Transit System Option, Model: Quickie LXI (EILXI), Product is manufactured and distributed by Sunrise Medical, Inc. Fresno, CA Intended to provide mobility to persons restricted to a sitting position and are unable to transfer from their wheelchairs when traveling in motor vehicles, the wheelchair must serve as a vehicle seat.
Brand
Sunrise Medical Inc
Lot Codes / Batch Numbers
Serial Numbers: LXI-034554, LXI-034697, LXI-034694, LXI-034500, LXI-034609, LXI-034740, and LXI-034741.
Products Sold
Serial Numbers: LXI-034554, LXI-034697, LXI-034694, LXI-034500, LXI-034609, LXI-034740, and LXI-034741.
Sunrise Medical Inc is recalling Manual Wheelchairs with Transit System Option, Model: Quickie LXI (EILXI), Product is manufactured due to The stabilizer bar was inadvertently omitted during the manufacturing process.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The stabilizer bar was inadvertently omitted during the manufacturing process.
Recommended Action
Per FDA guidance
An Urgent: Medical Device Correction letter, dated February 2009, with certified return receipt was sent to customers in the United States; Canadian consignees were notified via e-mail. The letter identified the affected product and the reason for the recall. The letters asked customers to contact their Quickie dealer to make arrangements to have their device wheelchair corrected at no charge and to also verify their wheelchair serial number on the form. Questions should be directed towards their dealer or Sunrise Medical at 800-456-8168.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026