Tandem Diabetes Care Inc t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology
Brand
Tandem Diabetes Care Inc
Lot Codes / Batch Numbers
Dexcom G5 - All Serial Numbers, Basal-IQ Technology - Software Versions 6.3.01 and Earlier
Products Sold
Dexcom G5 - All Serial Numbers; Basal-IQ Technology - Software Versions 6.3.01 and Earlier
Tandem Diabetes Care Inc is recalling t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technolog due to Insulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Insulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the battery life drops to less than 5%, insulin delivery will continue for 30 minutes and then the pump will power off and insulin delivery will stop. If the battery reaches 1%, then insulin delivery will stop, which could lead to hyperglycemia.
Recommended Action
Per FDA guidance
On 09/10/2020, correction notices were sent to customers who were asked to acknowledge Low Power Alerts and Low Power Alarms by tapping "OK" and then charging the pump. Further, take the following precautions to ensure that your pump is properly charged, and the pump's battery life does not fall below 25% remaining power: - Periodically check the battery level indicator of your pump - Charge the pump for a short period of time every day (10 to 15 minutes) - Avoid full battery discharges A software update is planned by the end of 2020 that resolves the potential fluctuating battery life display. We will reach out to you to complete the free remote update of your insulin pump when the software update becomes available. Complete a online form to confirm that you have received this important user communication. Customers with questions were encouraged to contact diabetes@amsl.com.au In November 2020 software version 6.4.1 was released for all affected insulin pumps except Dexcom G5 Mobile CGM system pumps. On 12/06/2020, Dexcom G5 Mobile CGM system customers were informed that a new software update was available to address the battery issues but that the update requires using the Dexcom G6 CGM system. On 5/24/22, correction notices were sent to customers in Australia to update to the latest software. t:slim X2 insulin pump with Basal-IQ technology version 6.6 and Control-IQ technology version 7.6 are available with the latest software versions. Software update contact: AMSL: diabetes@amsl.com.au or 1300 851 056.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution