Tecan U S Inc Disposable Tips 1000 l with filter Conductive disposable tips for Tecan Genesis Series * Order Number 612513.1 * Tecan Schweiz AG Seestrae 103, CH-8708 Mannedorf, Switzerland * Made in Germany. Blister 1 x 96 pcs. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Disposable Tips 1000 l with filter Conductive disposable tips for Tecan Genesis Series * Order Number 612513.1 * Tecan Schweiz AG Seestrae 103, CH-8708 Mannedorf, Switzerland * Made in Germany. Blister 1 x 96 pcs.
Brand
Tecan U S Inc
Lot Codes / Batch Numbers
Part Number 10612523, Lot No. U122098L-1325, Lot No. U122098L-3225 & Lot No. U122098L-2325. The sub-lot numbers (i.e. 1325
Products Sold
Part Number 10612523, Lot No. U122098L-1325, Lot No. U122098L-3225 & Lot No. U122098L-2325. The sub-lot numbers (i.e. 1325, 3225 and 2325) are printed on each blister and stamped in blue colour on the shipping container in the upper right corner of the box.
Tecan U S Inc is recalling Disposable Tips 1000 l with filter Conductive disposable tips for Tecan Genesis Series * Order Numbe due to Double filtering of 1000 l tips. The defect could result in cross contamination of other samples due to possible dripping of sample while moving over. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Double filtering of 1000 l tips. The defect could result in cross contamination of other samples due to possible dripping of sample while moving over other samples.
Recommended Action
Per FDA guidance
The consignee was notified via fax''d letter on/about 11/08/2005.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
IL