Teleflex Medical Sheridan Preformed Uncuffed Oral Tracheal Tube, 3.0 mm; a Rx, sterile, single use tracheal tube; Hudson Respiratory Care Inc., Temecula, CA 92589-9020, Manufactured in Mexico; catalog number (REF) 5-22006 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Sheridan Preformed Uncuffed Oral Tracheal Tube, 3.0 mm; a Rx, sterile, single use tracheal tube; Hudson Respiratory Care Inc., Temecula, CA 92589-9020, Manufactured in Mexico; catalog number (REF) 5-22006
Brand
Teleflex Medical
Lot Codes / Batch Numbers
catalog number (REF) 5-22006, lot number 1189038
Products Sold
catalog number (REF) 5-22006, lot number 1189038
Teleflex Medical is recalling Sheridan Preformed Uncuffed Oral Tracheal Tube, 3.0 mm; a Rx, sterile, single use tracheal tube; Hud due to Biological indicator failed. Investigation eliminated all equipment and product related factors.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Biological indicator failed. Investigation eliminated all equipment and product related factors.
Recommended Action
Per FDA guidance
Teleflex Medical sent Urgent Medical Device Recall letters dated 9/27/06 via UPS overnight to the direct accounts, informing them that the product did not pass validated sterility test requirements, and requesting that they cease use of the affected units and return them to Teleflex Medical for replacement. If the accounts further distributed the product, they were requested to conduct a sub-recall of the affected units from their customers. The accounts were instructed to complete and return by fax the enclosed Acknowledgement & Stock Status Form, indicating the number of units being returned for replacement. Any questions were directed to Linda Todd at 1-800-334-9751, ext. 4951.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026