Terumo Cardiovascular Systems Corp Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.
Brand
Terumo Cardiovascular Systems Corp
Lot Codes / Batch Numbers
Serial numbers 0011 through 0039 and 0041 through 0500.
Products Sold
Serial numbers 0011 through 0039 and 0041 through 0500.
Terumo Cardiovascular Systems Corp is recalling Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source r due to Pump may stop without an audible tone as a result of quickly clearing the internal pump alarm condition or overpressure alert/alarm condition.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Pump may stop without an audible tone as a result of quickly clearing the internal pump alarm condition or overpressure alert/alarm condition.
Recommended Action
Per FDA guidance
Consignees were notified of the problem via letter dated 11/10/06 and informed that they would be contacted to schedule a service call to install software upgrades. Installation of software upgrades began at customer locations on 10/16/06.