Terumo Cardiovascular Systems Corp Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.
Brand
Terumo Cardiovascular Systems Corp
Lot Codes / Batch Numbers
Serial numbers 0011 through 0395 may be affected.
Products Sold
Serial numbers 0011 through 0395 may be affected.
Terumo Cardiovascular Systems Corp is recalling Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source r due to The power supply may fail to function due to a defective capacitor. (Note; this device incorporates a redundant separate power supply not affected by. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The power supply may fail to function due to a defective capacitor. (Note; this device incorporates a redundant separate power supply not affected by this capacitor)
Recommended Action
Per FDA guidance
The power supplies were replaced as complaints were received between August 2005 and February 2006. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their units will be corrected when replacement components become available.