Terumo Cardiovascular Systems Corporation Terumo Sarns Level Sensor II Pads. For use with Sarns Advanced Perfusion System I, Perfusion System 800 and Perfusion System 900. Terumo Cardiovascular Systems Corp., Ann Arbor, MI.; REF 195240. Level Sensor pads are used to attach the level sensor to the sides of a rigid reservoir. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Terumo Sarns Level Sensor II Pads. For use with Sarns Advanced Perfusion System I, Perfusion System 800 and Perfusion System 900. Terumo Cardiovascular Systems Corp., Ann Arbor, MI.; REF 195240. Level Sensor pads are used to attach the level sensor to the sides of a rigid reservoir.
Brand
Terumo Cardiovascular Systems Corporation
Lot Codes / Batch Numbers
Lot Number 78925.
Products Sold
Lot Number 78925.
Terumo Cardiovascular Systems Corporation is recalling Terumo Sarns Level Sensor II Pads. For use with Sarns Advanced Perfusion System I, Perfusion System due to The level sensor holder may detach from the adhesive pad causing the sensor to lose contact with the venous reservoir, which may result in a detached . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The level sensor holder may detach from the adhesive pad causing the sensor to lose contact with the venous reservoir, which may result in a detached sensor indication or false alarm.
Recommended Action
Per FDA guidance
Consignees were notified via recall letter (Urgent Medical Device Removal letter) dated 2/13/08 and requested to report stocks on hand so that replacements could be sent.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026