Terumo Cardiovascular Systems Corporation Terumo TenderFlow Pediatric Arterial Cannulae, 2.0 mm (6Fr) OD, with 1/4 inch vented connector, 9 inch (22.9 cm) long; Catalog no. 813567. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Terumo TenderFlow Pediatric Arterial Cannulae, 2.0 mm (6Fr) OD, with 1/4 inch vented connector, 9 inch (22.9 cm) long; Catalog no. 813567. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
Brand
Terumo Cardiovascular Systems Corporation
Lot Codes / Batch Numbers
Lots 0501557 and 0512008.
Products Sold
Lots 0501557 and 0512008.
Terumo Cardiovascular Systems Corporation is recalling Terumo TenderFlow Pediatric Arterial Cannulae, 2.0 mm (6Fr) OD, with 1/4 inch vented connector, 9 in due to The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and repla. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The introducer or stylet is difficult to remove from the pediatric arterial cannula body, which may result in the need to remove the cannula and replace it.
Recommended Action
Per FDA guidance
An Urgent Medical Device Recall letter dated 10/16/08 was sent to consignees advising them of the problem; requesting that they immediately verify whether or not they had unused cannulae with the affected lot numbers; requesting that they isolate any affected product to prevent use; complete the Customer Response Form, indicate on form if it is medically necessary to use product without introducer; fax Customer Response Form to Terumo CVS at the number listed; and Terumo CVS would contact customers to arrange for the return of unused product. The letter stated that credit would be issued for any unused TenderFlow cannulae that were returned. The quantities to be returned should be indicated on the Customer Response Form. Consignees were advised that, if it was medically necessary to use the affected cannulae, they should be used without the introducer and in accordance with instructions for such use.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026