Trek Diagnostic Systems HARDY DIAGNOSTIC McFarland Standard 0.5; in vitro diagnostic, glass vial, one vial per box Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
HARDY DIAGNOSTIC McFarland Standard 0.5; in vitro diagnostic, glass vial, one vial per box
Brand
Trek Diagnostic Systems
Lot Codes / Batch Numbers
Model Number: Hardy = MSO5, Lot number 08182, Expiration date: 10/28/08
Products Sold
Model Number: Hardy = MSO5; Lot number 08182; Expiration date: 10/28/08
Trek Diagnostic Systems is recalling HARDY DIAGNOSTIC McFarland Standard 0.5; in vitro diagnostic, glass vial, one vial per box due to Incorrect result: The absorbance has changed as the standards have aged; the standards' turbidity has become denser. The recalled lot is now reading . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Incorrect result: The absorbance has changed as the standards have aged; the standards' turbidity has become denser. The recalled lot is now reading 0.15 ABS @ 625 nm. TREK'S internal specification and the manufacturer's (Hardy Diagnostics) specification is 0.08 - 0.10 ABS @625 nm. CLSI specifies 0.08 - 0.13 ABS @ 625 nm.
Recommended Action
Per FDA guidance
On 9/22/2008 TREK Diagnostic sent a recall notification letter to their customers advising them of the problem with the product. They instructed their customers to inventory their stock for any remaining recall lot and immediately destroy.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
IL, IN, OH