Trionix Research Laboratory Inc BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System.
Brand
Trionix Research Laboratory Inc
Lot Codes / Batch Numbers
Serial numbers B-102 through B-184
Products Sold
Serial numbers B-102 through B-184
Trionix Research Laboratory Inc is recalling BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging due to The ball screw on the unit which supports the detector head (which weighs approximately 800 lbs.) could snap resulting in the detector head dropping o. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The ball screw on the unit which supports the detector head (which weighs approximately 800 lbs.) could snap resulting in the detector head dropping on the patient.
Recommended Action
Per FDA guidance
The recalling firm modified one of the devices in 2002. The remaining device users were telephoned on/about 11/02/2006. To date the recalling firm has not taken any corrective action regarding devices with serial numbers B-102, B-103, B-104 or B-106 thru B-184.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026