Valleylab, Inc Radionics Cool-Tip RF Electrode Kit, containing DGP-HP single-use grounding pads. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Radionics Cool-Tip RF Electrode Kit, containing DGP-HP single-use grounding pads.
Brand
Valleylab, Inc
Lot Codes / Batch Numbers
Catalog Numbers: CT-1020, CT-1030, CT-1510, CT-1520, CT-1530, CT-2020, CT-2030, CT-2530, CTC-1025, CTC-1525, CTC-2025.
Products Sold
Catalog Numbers: CT-1020, CT-1030, CT-1510, CT-1520, CT-1530, CT-2020, CT-2030, CT-2530, CTC-1025, CTC-1525, CTC-2025.
Valleylab, Inc is recalling Radionics Cool-Tip RF Electrode Kit, containing DGP-HP single-use grounding pads. due to Pads may have a thin plastic film covering the conductive gel surface underneath the normal printed cover, causing non-adherence and possible burn.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Pads may have a thin plastic film covering the conductive gel surface underneath the normal printed cover, causing non-adherence and possible burn.
Recommended Action
Per FDA guidance
Users were notified by letter on 7/16/2003.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026