Varian Medical Systems Inc 4D Integrated Treatment Console and Varis 1.4g Medical Charged Particle Radiation Therapy System; All 4D Integrated treatment console versions 8.0.15 and below (excluding v7.0.31) and all Varis Non-Varian Treatment versions 6.6.5027 and below, with Serial numbers H510946, H510651, H510190. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
4D Integrated Treatment Console and Varis 1.4g Medical Charged Particle Radiation Therapy System; All 4D Integrated treatment console versions 8.0.15 and below (excluding v7.0.31) and all Varis Non-Varian Treatment versions 6.6.5027 and below, with Serial numbers H510946, H510651, H510190.
Brand
Varian Medical Systems Inc
Lot Codes / Batch Numbers
with Serial numbers H510946, H510651, H510190
Products Sold
All 4D Integrated treatment console versions 8.0.15 and below (excluding v7.0.31) and all Varis Non-Varian Treatment versions 6.6.5027 and below, with Serial numbers H510946, H510651, H510190
Varian Medical Systems Inc is recalling 4D Integrated Treatment Console and Varis 1.4g Medical Charged Particle Radiation Therapy System; A due to Selecting an empty space next to the ''Override'' or ''Acquire Actual'' buttons may result in mistreatment to the patient.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Selecting an empty space next to the ''Override'' or ''Acquire Actual'' buttons may result in mistreatment to the patient.
Recommended Action
Per FDA guidance
On September 21, 2006, the firm issued letters via certified mailing with return receipt to all its direct consignees, informing them of the affected product and providing instructions on the recall.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.