Varian Medical Systems Inc Varian brand VARiS 1.4g (Medical Charged Particle Radiation Therapy System, Record and Verify System); Linear Accelerator with RTP Exchange v6.2, v6.6, & v8.0, All Models; Varian Medical Systems, Palo Alto, CA Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Varian brand VARiS 1.4g (Medical Charged Particle Radiation Therapy System, Record and Verify System); Linear Accelerator with RTP Exchange v6.2, v6.6, & v8.0, All Models; Varian Medical Systems, Palo Alto, CA
Brand
Varian Medical Systems Inc
Lot Codes / Batch Numbers
Serial Numbers: All
Products Sold
Serial Numbers: All
Varian Medical Systems Inc is recalling Varian brand VARiS 1.4g (Medical Charged Particle Radiation Therapy System, Record and Verify System due to Incorrect Software Validation- The ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery that will not be recognized properly by the sys. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Incorrect Software Validation- The ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery that will not be recognized properly by the system. Under these circumstances, an arc plan can be administered to a patient whereby the gantry will not rotate, resulting in delivery to a single location rather than through the gantry arc.
Recommended Action
Per FDA guidance
An Urgent Medical Device Correction letter was distributed October 2, 2007, to all direct consignees, informing them that the ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery will not be recognized properly by the System. The firm provided corrective instructions and informed consignees that a product modification will be developed to resolve the issue. A Varian Service Representative will be scheduling a visit to upgrade the application.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026