Verathon, Inc. GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngoscope System Per PLRA: The Core OneTouch Cables (Reusable Device) is ancillary equipment used with the Core 10 and Core 15 monitors. The monitor and workstation are intended to work with video endoscopes, in conjunction with ancillary equipment, for endoscopic procedures. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngoscope System Per PLRA: The Core OneTouch Cables (Reusable Device) is ancillary equipment used with the Core 10 and Core 15 monitors. The monitor and workstation are intended to work with video endoscopes, in conjunction with ancillary equipment, for endoscopic procedures.
Brand
Verathon, Inc.
Lot Codes / Batch Numbers
All serial numbers
Products Sold
All serial numbers
Verathon, Inc. is recalling GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngosc due to When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port,. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.
Recommended Action
Per FDA guidance
On 06/05/20, Field Safety Notice/ URGENT:MEDICAL DEVICE RECALL notices were sent to customers. Customers were asked to do the following: - Review the information contained within this Field Safety Notice with all staff members who should be aware of the contents of the Field Safety Notice. - Discontinue use of all affected cables. - When replacements are received, destroy or return all affected cables in your possession. - Complete and return the recall response form. Customers with additional questions are encouraged to contact Customer Care agent support Monday through Friday from 6:00am to 4:00pm Pacific Time for the United States at (800) 331-2313 or email CSNotifications@verathon.com