Viasys Respiratory Care, Inc. Viasys V.I.P. Bird Gold pediatric ventilator, Catalog Number 15653 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Viasys V.I.P. Bird Gold pediatric ventilator, Catalog Number 15653
Brand
Viasys Respiratory Care, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Serial numbers: AAM01001 to AMMO1590 + KJI01178 for Domestic Customers and AAMO1648 to AEM01298 for International Customers
Viasys Respiratory Care, Inc. is recalling Viasys V.I.P. Bird Gold pediatric ventilator, Catalog Number 15653 due to The reason for this correction is due to the potential of a repeat ''failure to cycle'' event.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The reason for this correction is due to the potential of a repeat ''failure to cycle'' event.
Recommended Action
Per FDA guidance
On November 2, 2006, a product notification letter was mailed to all customers in the United States via overnight carrier (Federal Express). Consignees are informed in the notification letter that service calls are in process of being scheduled to update domestic consignees. On November 4, 2006, a product notification letter was mailed to all international customers via registed mail. Consignees are informed in the notification letter that replacement software, replacement instructions, and installation verification forms are included in their recall package.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026