Vitrolife Inc SpermMar Test 0.7ml Beads Particles Label on bottle: SpermMar Test IgA 0.7ml Beads/Particles-50 Tests IND REF SPMA_S LOT Expiration DO NOT FREEZE/NE PAS CONGELER FertiPro N.V. Label on box: SpermMar Test IGA 50 Determination Test Kit CONTENTS 1 vial SpermMar Latex Particles 0.7 ml REF SPMA_S FertiPro FertiPro N.V. Label on box: SpermMar Test IgA 50 Determination Test Kit Contents: 1 vial SpermMar latex particles, 0.7ml FertiPro N.V. For In Vitro Diagnostic Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SpermMar Test 0.7ml Beads Particles Label on bottle: SpermMar Test IgA 0.7ml Beads/Particles-50 Tests IND REF SPMA_S LOT Expiration DO NOT FREEZE/NE PAS CONGELER FertiPro N.V. Label on box: SpermMar Test IGA 50 Determination Test Kit CONTENTS 1 vial SpermMar Latex Particles 0.7 ml REF SPMA_S FertiPro FertiPro N.V. Label on box: SpermMar Test IgA 50 Determination Test Kit Contents: 1 vial SpermMar latex particles, 0.7ml FertiPro N.V. For In Vitro Diagnostic
Brand
Vitrolife Inc
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Manufacturer's Product Number/Catalog Number: SPMA_S Lot/Serial Number Expiration Date FP14A09 02/29/2016 FP14A10 04/30/2016 FP14A11 04/30/2016
Vitrolife Inc is recalling SpermMar Test 0.7ml Beads Particles Label on bottle: SpermMar Test IgA 0.7ml Beads/Particles-5 due to The firm is recalling three batches (FP14A09, FP14A10, FP14A11) of SpermMAR IgA test due to new formulation causing the motility of sperm cells to red. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The firm is recalling three batches (FP14A09, FP14A10, FP14A11) of SpermMAR IgA test due to new formulation causing the motility of sperm cells to reduce in certain samples when the new bead solution was added.
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
OR
Page updated: Jan 11, 2026