W L Gore & Associates, Inc. Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diameter 16 mm, UDI: Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diameter 16 mm, UDI:
Brand
W L Gore & Associates, Inc.
Lot Codes / Batch Numbers
Serial Numbers: 22753580
Products Sold
Serial Numbers: 22753580
W L Gore & Associates, Inc. is recalling Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LA due to Incorrect labeling, specifically, expandable endoprosthesis labeled as 8 mm x 59mm x 135 cm, however may contain a 9mm x 29 mm x 135 cm device.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Incorrect labeling, specifically, expandable endoprosthesis labeled as 8 mm x 59mm x 135 cm, however may contain a 9mm x 29 mm x 135 cm device.
Recommended Action
Per FDA guidance
On September 9, 2021, Gore issued a "Urgent Field Safety Notice" to affected consignees via E-Mail. In addition to informing consignees about the recalled product, Gore asked consignees to take the following actions: 1. Identify and return the device listed in the CHANNEL PARTNER RESPONSE FORM if unused. 2. Please complete and sign the enclosed CHANNEL PARTNER RESPONSE FORM and return to FieldActionTeam@wlgore.com within 2 weeks of receipt of this notification. 3. Please share this letter with others in your hospital or clinic as appropriate. 4. If a listed device has been implanted, there is no patient-follow up needed and there are no further actions needed other than informing Gore the device was used. Please indicate the used device on the CHANNEL PARTNER RESPONSE FORM and return to FieldActionTeam@wlgore.com within 2 weeks of receipt of this letter. No further action is needed. 5. Any adverse event involving the GORE VIABAHN VBX Balloon Expandable Endoprosthesis should be reported to the manufacturer and the country specific regulatory authorities immediately. To report an event to W. L. Gore & Associates, email: medcomplaints@wlgore.com or contact: EMEA: +49 89 4612 3440, Fax: +49 89 4612 43440 6. This notice needs to be passed on to all those who need to be aware within your institution or to any organization where potentially affected devices have been transferred (as appropriate). Please transfer this notice to other organization(s) on which this action has an impact (as appropriate).
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.