Surveyor S12/S19 Monitor (Welch Allyn) – Defibrillation Risk (2020)
Medical device may not meet defibrillation withstand requirements.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA SNAP JSCREW, PAT CBL 10WIRE IEC SNAP JSCREW. The Surveyor S12 patient monitor is designed for continuous monitoring in either stationary or portable applications.
Brand
WELCH ALLYN, INC/MORTARA
Lot Codes / Batch Numbers
Model Numbers: SUR12-XXX-XXXXX, SUR19-XXX-XXXXX, 9293-033-52, 9293-033-53.
Products Sold
Model Numbers: SUR12-XXX-XXXXX, SUR19-XXX-XXXXX, 9293-033-52, 9293-033-53.
WELCH ALLYN, INC/MORTARA is recalling Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: due to Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Recommended Action
Per FDA guidance
Welch Allyn notified customers on 2/19/2020 via "Voluntary Medical Device Correction Welch Allyn Patient Cables or Lead Sets". The recall letter identified the affected product and stated the reason for the recall. The firm also requested the distributors to immediately notify accounts if the product was further distributed.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026