Welch Allyn Protocol, Inc Welch Allyn Vital Signs Monitor 5300 Series 300 Part numbers 007-0098-00 through 007-0159-00 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Welch Allyn Vital Signs Monitor 5300 Series 300 Part numbers 007-0098-00 through 007-0159-00
Brand
Welch Allyn Protocol, Inc
Lot Codes / Batch Numbers
Serial numbers JA000225 through JA002052.
Products Sold
Serial numbers JA000225 through JA002052.
Welch Allyn Protocol, Inc is recalling Welch Allyn Vital Signs Monitor 5300 Series 300 Part numbers 007-0098-00 through 007-0159-00 due to Potential speaker failure could result in absence of audible annunciation of pulse tones, patient alarms and equipment alerts.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential speaker failure could result in absence of audible annunciation of pulse tones, patient alarms and equipment alerts.
Recommended Action
Per FDA guidance
The firm contacted their customers and their distributors customers via letters dated June 8, 2004 through July 9, 2004. The letters instruct their customers to verify speaker functionality. If the customers do not hear two quick beeps or there is static then they have a faulty speaker. The letters instruct them to stop using the monitor and contact Welch Allyn to have the monitor repaired. If the monitor continues to emit two quick beeps their customers are to contact Welch Allyn to have the monitor repaired as soon as it is convenient. The customers are to verify the two beeps prior to each use. The letter dated June 8, 2004 went to only a couple of direct customers. The remaining letters dated June 14, 2004, June 23, 2004, June 30, 2004 and July 9, 2004 went to direct customers, distributors and distributors'' end user customers.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026