Western a Scott Fetzer Company Medical Gas Yoke Assembly. Airgas Portable 4-Cylinder Manifold Cart Model #AGI-2238. The device is a medical gas yoke assembly which is used to connect gas cylinder post valves to equipment used for anesthesia or respiratory therapy. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medical Gas Yoke Assembly. Airgas Portable 4-Cylinder Manifold Cart Model #AGI-2238. The device is a medical gas yoke assembly which is used to connect gas cylinder post valves to equipment used for anesthesia or respiratory therapy.
Brand
Western a Scott Fetzer Company
Lot Codes / Batch Numbers
Oxygen Yoke Assembly Model # 8065-870
Products Sold
Oxygen Yoke Assembly Model # 8065-870, Nitrous Oxide Yoke Assembly Model # 8066-910 Airgas Portable 4-Cylinder Manifold Cart Model #AGI-2238
Western a Scott Fetzer Company is recalling Medical Gas Yoke Assembly. Airgas Portable 4-Cylinder Manifold Cart Model #AGI-2238. The device is due to The yoke pins of the attached medical gas yoke assemblies could become displaced to such a degree that the yoke pin index safety system could become c. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The yoke pins of the attached medical gas yoke assemblies could become displaced to such a degree that the yoke pin index safety system could become compromised and allow a post valve of one gas type to become interchangeable with a yoke connection of another gas type.
Recommended Action
Per FDA guidance
Western Enterprises issued an "Urgent-Device Recall" notification dated August 20, 2009 informing Consignees of the affected product. Consignees are to contact the firm to request backup plates and installation instructions or request an RGA and shipping instructions for return of the manifold block assembly for modification. For further information, contact Western at 1-800-783-7890.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.