Wilson-Cook Medical Inc. Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7, REF G56572; b) HEMO-7-EU, REF G24663; Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7, REF G56572; b) HEMO-7-EU, REF G24663;
Brand
Wilson-Cook Medical Inc.
Lot Codes / Batch Numbers
UDI/DI : a) 00827002565722, b) 00827002246638, All unexpired lot numbers
Products Sold
UDI/DI : a) 00827002565722, b) 00827002246638; All unexpired lot numbers
Wilson-Cook Medical Inc. is recalling Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7, REF G56572; b) HEMO-7-EU, REF G2 due to There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.
Recommended Action
Per FDA guidance
The firm notified their consignees by letter and email beginning on 05/22/2023. The notice explained the issue, risk, and requested the following: Complete and return the Acknowledgement Form, ensure the notification is shared with appropriate personnel, down to the user level or customers if devices were further distributed, report any adverse events.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.