Wright Medical Technology, Inc. ORTHOLOC 3DSi CLAW II 4 HOLE Plate, Size 25MM, REF 402S0425, Bone fixation device Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ORTHOLOC 3DSi CLAW II 4 HOLE Plate, Size 25MM, REF 402S0425, Bone fixation device
Brand
Wright Medical Technology, Inc.
Lot Codes / Batch Numbers
GTIN 00840420152985, Lot Number 1636048
Products Sold
GTIN 00840420152985, Lot Number 1636048
Wright Medical Technology, Inc. is recalling ORTHOLOC 3DSi CLAW II 4 HOLE Plate, Size 25MM, REF 402S0425, Bone fixation device due to Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled,. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting in the incorrect product in packages. Packaging for 402S0425 incorrectly contained 40241425, and vice versa.
Recommended Action
Per FDA guidance
The firm issued an URGENT MEDICAL DEVICE notice to its the sole consignee by email on 02/02/2023. The notice explained the product and the issue and requested the following actions be taken: 1. Check inventory for affected product and remove from use 2, Sign and return the enclosed Business Reply Form by email to fieldaction@stryker.com to confirm receipt of this notification/documenting product disposition. Stryker will will contact you to arrange for the return of the affected device(s) 3. Maintain awareness of this communication internally until all required actions have been completed within your facility.. 4. If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details so that we can inform the recipients appropriately. b. If you are a distributor, note that you are responsible for notifying your affected customers.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026