Xoran Technologies, Inc. Xoran MiniCAT CT Scanner ID ENT00001 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Xoran MiniCAT CT Scanner ID ENT00001
Brand
Xoran Technologies, Inc.
Lot Codes / Batch Numbers
minicat
Products Sold
minicat
Xoran Technologies, Inc. is recalling Xoran MiniCAT CT Scanner ID ENT00001 due to Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.
Recommended Action
Per FDA guidance
The service report in response to this complaint stated that the Head select relay board, chip and fan were replaced. The machine (ENTOOOO I) was subsequently tested. All tests were successful.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026