XTANT Medical Holdings, Inc Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length. Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.
Brand
XTANT Medical Holdings, Inc
Lot Codes / Batch Numbers
UDI-DI: 00810076564005, Lot: AP23000143
Products Sold
UDI-DI: 00810076564005, Lot: AP23000143
XTANT Medical Holdings, Inc is recalling Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter due to Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is inc. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.
Recommended Action
Per FDA guidance
On 11/28/23 the recall notice was emailed. Customers with questions can call the firm at 1-406-924-5878 from 8 a.m. to 5 p.m. (Mountain Time).