Zhejiang Kindly Medical Devices Co., Ltd. # 758 Binghai Five Rd. Longwan Dist Wenzhou China HENRY SCHEIN Safety Blood Collection Needle 21Gx 1 1/4 Inch, REF 570-2067, 100pcs/box 1000pcs/ctn 350*210*350mm Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
HENRY SCHEIN Safety Blood Collection Needle 21Gx 1 1/4 Inch, REF 570-2067, 100pcs/box 1000pcs/ctn 350*210*350mm
Brand
Zhejiang Kindly Medical Devices Co., Ltd. # 758 Binghai Five Rd. Longwan Dist Wenzhou China
Lot Codes / Batch Numbers
Lot Numbers / Expiration Date: 20201030 / Exp. Date 2025.10.29, 20201210 / Exp. Date 2025.12.09.
Products Sold
Lot Numbers / Expiration Date: 20201030 / Exp. Date 2025.10.29, 20201210 / Exp. Date 2025.12.09.
Zhejiang Kindly Medical Devices Co., Ltd. # 758 Binghai Five Rd. Longwan Dist Wenzhou China is recalling HENRY SCHEIN Safety Blood Collection Needle 21Gx 1 1/4 Inch, REF 570-2067, 100pcs/box 1000pcs/ctn 35 due to Potential for difficulty in safety mechanism activation.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential for difficulty in safety mechanism activation.
Recommended Action
Per FDA guidance
On 8/24/2021, Zhejiang Kindly Medical Devices Co., Ltd. issued Urgent Medical Device Safety Alert notices to two customers. On 9/17/2021, issued a second notice to customers dated 8/24/2021. Action to be Taken by User: 1 - Advise the user to lower the angle between safety protective cap and needle tube by slight pressing in the direction of activation, then using a thumb or hard surface to activate the safety protective cap and lock the needle tube in the IFU, then together with corresponding illustration. 2 - Request the user to strictly follow the guidance of the IFU.