Zimmer Inc. Zimmer brand Coonrad/Morrey total elbow interchangeable ulnar assembly, small, left, 3 in. length. tivanium TI-6AL-4V alloy/UHMWPE, sterile, for use with bone cement. Cat. no. 32-8105-53-01. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Zimmer brand Coonrad/Morrey total elbow interchangeable ulnar assembly, small, left, 3 in. length. tivanium TI-6AL-4V alloy/UHMWPE, sterile, for use with bone cement. Cat. no. 32-8105-53-01.
Brand
Zimmer Inc.
Lot Codes / Batch Numbers
Lot 60039452.
Products Sold
Lot 60039452.
Zimmer Inc. is recalling Zimmer brand Coonrad/Morrey total elbow interchangeable ulnar assembly, small, left, 3 in. length. t due to Right ulna implants were etched and labeled as left ulna implants. Left ulna implants may have been etched and labeled as right ulna implants.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Right ulna implants were etched and labeled as left ulna implants. Left ulna implants may have been etched and labeled as right ulna implants.
Recommended Action
Per FDA guidance
Consignees were notified via e-mail on 3/16/04. Consignees were asked to secure their inventory from use and to notify all customers if they had further distributed the product. Product shipped to Japan and Singapore was contained prior to shipment to hospitals.