Zimmer Inc. Zimmer Coonrad/Morrey Total Elbow, Interchangeable Humeral Assembly, extra small, 4 inch length, sterile, Zimmer Inc., Warsaw, IN; REF 32-8105-027-04. The device is utilized for a total elbow arthroplasty surgical procedure. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Zimmer Coonrad/Morrey Total Elbow, Interchangeable Humeral Assembly, extra small, 4 inch length, sterile, Zimmer Inc., Warsaw, IN; REF 32-8105-027-04. The device is utilized for a total elbow arthroplasty surgical procedure.
Brand
Zimmer Inc.
Lot Codes / Batch Numbers
All lots.
Products Sold
All lots.
Zimmer Inc. is recalling Zimmer Coonrad/Morrey Total Elbow, Interchangeable Humeral Assembly, extra small, 4 inch length, ste due to The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.
Recommended Action
Per FDA guidance
Consignees were notified by "Urgent Device Correction" letter dated January 30, 2009 and provided revised surgical techniques for these extra-small implants. Direct questions about this recall to Zimmer, Inc. at 574-267-6131.