Zimmer Manufacturing B.V. Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni, multi, and cluster-holed. The multi and cluster-holed shells allow for the use of screws to provide additional fixation and security, particularly in those case where acetabular bone stock is deficient. The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninfla Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni, multi, and cluster-holed. The multi and cluster-holed shells allow for the use of screws to provide additional fixation and security, particularly in those case where acetabular bone stock is deficient. The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninfla
Brand
Zimmer Manufacturing B.V.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Item Number Lot Number 00-6250-065-35 62748089 00-6250-065-35 62754329 00-6250-065-35 62793494 00-6250-065-35 62793495 00-6250-065-35 62793501 00-6250-065-35 62793502 00-6250-065-35 62793503 00-6250-065-35 62813612 00-6250-065-35 62813613 00-6250-065-25 62784617 00-6250-065-25 62784618 2 00-6250-065-25 62784619 00-6250-065-25 62784621 00-6250-065-25 62784622 00-6250-065-25 62818701 00-6250-065-25 62818702 00-6250-065-25 62818703 00-6250-065-25 62818707 00-6250-065-25 62818709 00-6250-065-25 62818710 00-6250-065-25 62825611 00-6250-065-25 62836984
Zimmer Manufacturing B.V. is recalling Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni, multi, and due to Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific mach. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of the fractures. .
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 11, 2026