Zoll Medical Corporation Zoll M Series External Defibrillator Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Zoll M Series External Defibrillator
Brand
Zoll Medical Corporation
Lot Codes / Batch Numbers
Serial Numbers: T02I37097-T03C45238
Products Sold
Serial Numbers: T02I37097-T03C45238
Zoll Medical Corporation is recalling Zoll M Series External Defibrillator due to Defibrillator may fail to charge or hold its charge above energies of 75 joules. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Defibrillator may fail to charge or hold its charge above energies of 75 joules
Recommended Action
Per FDA guidance
Zoll Medical issued a letter on 6/25/03 to domestic consignees and 6/26/03 to international consignees on 6/26/03. Letters were issued Certified Mail Return Receipt requested. Instructions are provided for Biomedical to charge device and then dump the charge. A damaged device will fail to charge and issue an error message. Users then are requested to return device. Firm will initiate telephone contact on 6/30/03 to all accounts to verify receipt of notice and confirmation of testing per the instructions.