GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.