Patient may experience minor burn or blistering if device is exposed to direct sunlight and/or warm temperatures for over 12 hours or battery is damaged.
Patient may experience minor burn or blistering if device is exposed to direct sunlight and/or warm temperatures for over 12 hours or battery is damaged.
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.