Medtronic Model 3777-75 Octad 1x8 Leads, Lot V030733 contain accessory stylets that are not the correct length.
The Smith Medical PM, Inc. BCI 3180 Pulse Oximeter tantalum capacitors C38, C72, C74, and C98 were installed backwards during assembly at the board manufacturer. This exposes the components to reverse polarity voltages that may lead to their pre-mature failure. As a result, the identified monitors may exhibit failure modes such as the monitor may shut down and not produce alarms or signals to
Stryker Trauma has become aware that five lots of the Gamma3 Set Screwdriver have hexagon tips that were not hardened to specification. If the screwdriver tip were to deform intra-operatively, the correct positioning of the set screw would not be possible, which could lead to lateral or medial migration of the lag-screw.