Foreign material- this device, a sterile wound care product, was found to be contaminated with clumps of brown/black foreign material
Firm received complaints for signal 2 & 4 errors, and shifts in Relative Light Units (RLU). Investigations showed that the On-Board Stability (OBS) of the acid & base reagents may be compromised after 4 days after installation onto the Centaur CP system due to evaporation.
Following installation of software version 1.30 in October 2006, the firm confirmed a complaint that the new software caused a discrepancy in the pump speed calculation and caused the pump to stop during clinical use and a ''connect motor'' message to be displayed.